An comparison of a leading fully-automated ECG waveform component algorithm vs. computer-assisted expert analysis. Barbey, JT, White, KV, Pezzullo, JC, Affrime, M. Man vs. Machine: Are Cardiac Core Labs still Relevant? (2011). Journal of Clinical Pharmacology, 51:1343.
Kenn White
December 19, 2013
Annual Pro sale · Save $30
Have presentations of your own?
Share with private links, password protection, and more control over your work.
Transcript
9 slides
-
Barbey, JT, White, KV, Pezzullo, JC, Affrime, M. (2011) Man vs. Machine: Are Cardiac Core Labs still Relevant?. American College of Clinical Pharmacology Annual Meeting. J Clin Pharmacol, 51:1343. MAN vs. MACHINE: ARE CARDIAC CORE LABS STILL RELEVANT? J T Barbey,1 K V White,1 J C Pezzullo,1 M Affrime,2 1Social & Scientific Systems, Inc., Silver Spring, MD 2Icon Development Solutions, Ellicott City, MD ABSTRACT: Background: Most ECGs are recorded using digital equipment that can provide an
automated assessment of heart‐rate corrected (QTc) interval. Several manufacturers have developed new enhanced QTc algorithms but core lab‐read ECGs are still preferred in the context of thorough QT studies. The objective of this study was to compare the unedited Fridericia‐corrected QT (QTcF) values generated by Mortara’s VERITAS algorithms with those generated by a core lab‐based, computer assisted, manual measurement method (HeartSignalsTM, Social & Scientific Systems Inc) using a single over‐reading cardiologist. Methods: ECGs were acquired using the Surveyor telemetry central system (Mortara Instruments) during a prospective, single‐blind, placebo‐controlled, randomized, crossover study of moxifloxacin in 24 healthy volunteers. Maximum mean placebo‐corrected change from baseline QTcF (∆∆QTcF) was estimated by repeated measures analysis of averaged replicates for each method. Results: ECGs from 23 completed subjects at 3 pre‐dose and 12 post‐dose time points during the 2 treatment periods resulted in 2028 analyzable tracings, which were assessed independently by each method. Adequacy of the study sensitivity (lower bound of the 95% confidence interval of ∆∆QTcF >5 ms) was confirmed by the manual method at 3/12 and by the VERITAS algorithms at 7/12 post‐dose time points. The maximal mean ∆∆QTcF and 90% CIs were 12.3 (9.1‐15.5) and 18.1 (12.6‐23.7) ms respectively, occurring 3 hours post‐dose for both methods. Intra‐ triad QTcF variability, was 7.9 ms for the manual method and 9.7 ms for the VERITAS algorithms (P<0.001) The increased variability of the VERITAS measurements resulted from a number of excessively short QT measurements in several different subjects. Conclusion: In this population of normal volunteers, unedited automated QTcF values generated by the VERITAS algorithms showed greater variability and yielded a markedly larger moxifloxacin effect profile than those generated by a computer assisted manual method using a single over‐reading cardiologist -
INTRODUCTION: The techniques currently in use for the measurement of ECG intervals can be classified into three broad categories: fully manual, computer assisted with manual adjudication, and fully automated. The ICH E14 guidance
recommends either fully manual or computer assisted manual adjudication approaches for clinical trials in which the assessment of ECG safety is an important objective, such as the Thorough QT/QTc study. While concern remains that fully automated reading methods can yield misleading results in the presence of artifact, flat T waves or irregular rhythm, an increasing number of investigators have reported that the newer improved automatic algorithms can match and even surpass the precision of core lab manual or computer assisted measurements, particularly in the context of studies conducted normal volunteers.1 Our core lab recently participated in a prospective, single‐center, single‐blind, placebo‐ controlled, 2‐period, crossover study to assess the effect of moxifloxacin 400 mg on the QTc interval comparing the performance of our computer assisted manual approach (HeartSignalsTM, Social & Scientific Systems, Inc.) with that of an automated system (QTinno, NewCardio, Inc).2 Both the manual and automated approaches were successful in demonstrating assay sensitivity for a positive control and each demonstrated a high and comparable degree of precision. Since the Surveyor Telemetry Central System (Mortara Instrument, Inc) had been used to collect the raw 12‐lead ECG waveforms analyzed in the original study, we decided to compare our QTcF values to those generated by the Mortara VERITAS algorithm. METHODS: Study Design: This was a single-center, single-blind, placebo-controlled, 2-period, crossover study initially designed to assess the effect of moxifloxacin 400 mg on the QTc interval using our computer-assisted, manual approach with 100% cardiologist adjudication (HeartSignalsTM, Social & Scientific Systems, Inc.) and a fully automated system (QTinno, NewCardio, Inc). In the current study, we compared the QTcF values generated by HeartSignals with those generated by the VERITAS algorithm using the matching ECGs from our original analysis. Subjects participated in 2 treatment periods separated by a 1-week washout. Subjects received moxifloxacin 400 mg or placebo after an 8-hr fast. Subjects had baseline ECGs performed 5 times over 5 minutes at each of 3 pre-dose time points (0.75, 0.5, and 0.25 hr) and then 5 times at each of 12 time points from 0.5 hr through 24 hr post-dose. QT Measurements and Quality Control Process (QC): The Surveyor Telemetry Central System was used to collect the raw 12-lead ECG waveforms for analysis. To produce the original manual data set, a single cardiologist blinded to the treatment sequence and time order of the ECGs measured on a magnified screen the QT and RR of the 2nd through 4th replicate from each time point. Based on objective criteria, the 1st or 5th replicate was used in place of another, if fewer than 3 optimal replicates were -
reported after the initial read. In addition, the ECGs with
the longest QTcF values, the shortest QTcF values and a random sample of tracings with normal QTcF values were reviewed and corrected if needed. To produce the machine data set, QT, RR and QTcF values generated by the VERITAS algorithm were extracted from the exact tracings used for the manual analysis. The original unedited machine ECGs then underwent a QC process comparable to the one used for manually measured ECGs, generating a second “edited” machine QTcF dataset. No corrections were made to the VERITAS QT measurements but tracings with marking errors of >40 ms were excluded. The QTcF of each replicate was calculated from QT and RR. Intervals of each method were summarized by the arithmetic mean for data listings, summary, and analysis. Statistical Methods: The following analyses were performed separately using the unedited or edited, manual and machine generated QTcF data sets as indicated. To evaluate assay sensitivity, placebo corrected QTcF changes from baseline (ΔQTcF) were calculated using the mean arithmetic differences in QTcF. The arithmetic difference approach was preferred over the mixed model analysis of variance approach to allow for analysis of the unedited and edited datasets by a single approach. Assay sensitivity was considered adequate if the lower bound of at least one 95% one-sided confidence interval for the mean placebo-corrected ΔQTcF was greater than 5 msec and occurred at a biologically plausible time. An estimate of intra‐triad variability of single QTcF measurements was calculated for each data set and used to characterize precision of the respective ECG measurement approaches and data sets. RESULTS: QT data from 23 completed subjects at 30 time points during the 2 treatment periods resulted in 2028 valid, QC’d replicates by the manual approach. The initial unedited machine data set including the same 2028 ECGs showed significantly greater intra-triad variability than the manually measured data set (9.7 ms vs 7.9 ms, P<.001) This appeared due, in part at least, to implausibly short QTcF values as shown in figure 1. -
QC review of the machine read ECGs led to the
deletion of 116 of 127 ECGs with a QTcF<370 ms, 1 of 94 randomly chosen ECGs with a normal QTcF and 1 of 85 tracings with a QTcF >450 ms resulting in a new “edited” machine dataset with 1910 tracings. See figures 2a, 2b and 2c for examples of accepted and rejected measurements. FIGURE 2a. Accepted machine QT measurement (accurate) -
FIGURE 2b Rejected machine QT measurement (too short) FIGURE 2c
Rejected machine QT measurement (too long) As shown on figure 3, the QC process eliminated many of the excessively short QTcF values measured by the machine. Intra triad variability of the machine generated values after QC decreased from 9.7 to 8.3 msec but remained significantly greater than that of the matching manually read ECGs (8.3 msec vs. 7.8 msec, P<0.002). -
Although applied to a set of ECGs successfully read using
the manual approach, the unedited machine read QTcF values produced a suboptimal QTcF data set, characterized by unacceptably high intra-triad variability and several implausibly short QTcF measurements. Adequacy of the study sensitivity for the unedited data set(lower bound of the one-sided 95% confidence interval of ΔQTcF > 5 msec) was confirmed at only one time point using the arithmetic differences approach. The maximal mean ΔQTcF was 15.3 (6.5-24.1) msec and occurred at 3 hrs. (Figure 4a) Assay sensitivity was demonstrated at 7 time points using a mixed model of variance analysis, with positive values persisting beyond the expected decay curve for moxifloxacin. (Data not shown) The QC process produced an edited machine generated QTcF data set with fewer ECGs and lower intra triad variability. Adequacy of the study sensitivity for this edited data set (lower bound of the one-sided 95% confidence interval of ΔQTcF greater than 5 msec) was confirmed at 6 time point using the arithmetic differences approach. The maximal mean ΔQTcF was 16.7 (11.3-22.1)-24.1) msec and occurred at 3 hrs. (Figure 4b) Assay sensitivity was demonstrated at a 3 time points by the manual approach whether by arithmetic differences or mixed model of variance analysis. The maximal mean ∆QTcF by manual measurements was 12.2 (9.1-15.3) msec and occurred at 3 hrs as well. (Figure 4b) -
The summary of QTcF measured by machine (edited) vs. QTcF
measured by man differences suggests that intervals by the machine (edited) method were, on average, about 12 msec shorter than the intervals measured by man, with a standard deviation of +/- 15 msec. Mean ΔQTcF following moxifloxacin was typically 3.5 msec higher for edited machine reads than for man measured ECGs, while ΔQTcF following placebo was not systematically different between man and edited machine reads. DISCUSSION AND CONCLUSIONS: This study was designed to evaluate the performance of a commercially available automated ECG algorithm (VERITAS) when measuring QTcF on a set of good quality ECGs acquired in a population of normal volunteers receiving moxifloxacin and placebo in a crossover design. Compared with the values generated by a 100% manually adjudicated computer assisted system, the unedited data generated by the automated algorithm showed significantly greater intra triad QTcF variability due predominantly to erroneously short QT measurements. After completion of a QC process involving review of short and long QTcF outliers as well as a random sample of tracings with normal QTcF, a second edited machine-generated data set excluding markedly incorrectly measured tracings was produced. This second, smaller data set showed lower intra-triad variability still however greater than the matching man-generated data set. Although mean machine-measured QTcFs were shorter than those measured by the manual method, the machine-generated moxifloxacin effect profile was more pronounced than the profile -
based on manual reads, a difference that has been reported
with other automated reading systems1,2. In summary unedited automated QTcF values generated by the VERITAS algorithm showed greater intra triad variability and yielded a suboptimal moxifloxacin effect profile compared with the data generated from the same ECGs by a computer assisted, 100% manually adjudicated approach. Extensive QC and deletion of markedly incorrectly measured ECGs, increased the precision of the machine reads but still did not match the precision of the computer assisted 100% manually adjudicated system. These findings apply to the VERITAS automated algorithm and the HeartSignalsTM computer assisted 100% manually adjudicated approach and may not reflect the performance of other automated and computer assisted systems. REFERENCES: 1. Couderc J C, Garnett C, Li C, Handzel R, McNitt S, Xia X, Polonsky S, Zareba W. Highly Automated QT Measurement Techniques in 7 Thorough QT Studies Implemented under ICH E14 Guidelines Ann Noninvasive Electrocardiol 2011;16(1):13–24 2. Ruff D, Connolly M, Brueckner R, Bynum L, Beck D, Gussak I, Barbey J T, White K, Krantz M J, Affrime JM A prospective, single-blind, placebo-controlled, randomized, crossover study to assess the performance of automated and manual methodologies for detecting QTCc interval prolongation. Clin Pharmacol Ther . 2011; 89:s15